SPECIFICATIONS- MODELS TTU3 & TTU4C

 THE SIMUTESTERS MODEL TTU4C HAS TWO INDEPENDENT FREQUENCY CHANNELS, A MILLIAMP CHANNEL, A CHANNEL FOR MILLIVOLTS OR THERMOCOUPLE SIGNALS, A BUILT- IN MULTIMETER, AND VARIOUS OTHER  FEATURES.  MOST OUTPUTS ARE AVAILABLE SIMULTANEOUSLY, MAKING THE INSTRUMENT SUITABLE FOR SIMULATING COMPLEX CONTROL SYSTEMS.  THE INSTRUMENT IS HOUSED IN A PORTABLE CASE, WEATHER RESISTANT WHEN CLOSED.  DISPLAYS AND MULTIMETER ARE MOUNTED IN THE LID.  CONTROLS ARE MOUNTED IN THE BASE OF THE CASE. THE TTU3C IS IDENTICAL TO THE TTU4C WITH THE EXCEPTION THAT THE 3C HAS 1 FREQUENCY CHANNEL AND THE 4C HAS 2 FREQUENCY CHANNELS

These instrumemts are the standard of the industry and are used by most major organizations that service gas turbines.

 send email for more information

    FREQUENCY CHANNELS
         OUTPUT
                FREQUENCY RANGE                        1 HZ TO 25 KHZ
                OUTPUT VOLTAGE
                     SQUARE WAVE                             7.5 V P-P NO LOAD
                                                                              4 V P-P 300 OHM LOAD
                CONTROLLABILITY                          + / - 1 HZ
                ISOLATED FROM ALL OTHER CIRCUITS
                MONITORING - SPECS SAME AS FOR "INPUT"
            INPUT
                FREQUENCY RANGE                         1 HZ TO 50 KHZ
                SENSITIVITY                                       2.5 V P-P
                RESOLUTION                                      1 HZ
                ACCURACY                                         .02% OF READING + / - 1 HZ
                UPDATE RATE                                     1 PER SECOND
                ISOLATED FROM ALL OTHER CIRCUITS

    MILLIAMP CHANNEL
        OUTPUT
                RANGE                                                  0 TO 1 mA OR 0 TO 50 mA
                CONTROL ABILITY                             0.01 mA
                MAX RESISTIVE LOAD                      150 OHMS AT 20 mA
                LOAD REGULATION                          < 0.01 mA OUTPUT CHANGE FROM
                                                                               SHORT CIRCUIT TO 100 OHM LOAD
        INPUT
                READ WITH INTERNAL MULTIMETER
                RESOLUTION                                      0.01 mA
                ACCURACY
                 0 - 20 mA                                             0.1 % OF READING + / - 0.01 mA
                 >20 mA                                                0.4 % OF READING + / - 0.01 mA

    mV / THERMOCOUPLE SIMULATION CHANNEL
        OUTPUT
                mV RANGE                                          0 TO 50 mV
                TC SIMULATE RANGE
                TYPE K                                              AMBIENT TO 2295 DEG F
                CONTROLLABILITY                        0.01 mV OR 1 DEG F OR 1 DEG C
                DRIVE                                                10 mA MAX
                MONITORING
                RESOLUTION, ACCURACY  SAME AS FOR "INPUT"
        INPUT
                RANGE                                             0 TO 50 mV
                RESOLUTION                                  0.01 mV OR 1 DEG F OR 1 DEG C
                ACCURACY - mV                            0.1 % OF READING + / - 1 mV
                                       -TEMP                        0.1 % OF READING + / - 2 DEG F OR + / - 1 DEG C

 TOP

BUILT-IN MULTIMETER - MODIFIED FLUKE 75  ALL COMMON MULTIMETER FUNCTIONS, AUTO RANGING.  POWERED BY INSTRUMENT SUPPLY.  ELECTRICALLY ISOLATED FROM ALL OTHER CIRCUITS.  PERMANENTLY MOUNTED IN THE INSTRUMENT LID.  SOME ACCURACY'S ARE BETTER THAN STANDARD METER.  DC MILLIAMP ACCURACY IS AS GIVEN IN "MILLIAMP CHANNEL".

 INTERNAL BATTERY POWER
                5 V DC NI. CAD. 2.0 AH CAPACITY
                INSTRUMENT OPERATING TIME WITH FULLY CHARGED BATTERIES ---- 20 HOURS
                CHARGE TIME
                    FROM 110 V AC                         12 HRS
                    FROM 24 V DC                           12 HRS

    OPERATION FROM 110 V AC OR 24 V DC - THE INSTRUMENT MAY BE OPERATED FROM 110 V AC OR     24 V DC INDEFINITELY WITH NO PERFORMANCE DEGRADATION AND NO DAMAGE, WITH THE               BATTERIES IN ANY CONDITION.

LOW BATTERY INDICATION - AUDIBLE TONE INDICATES BATTERY VOLTAGE IS LESS THAN 4.8 V.  BATTERIES SHOULD BE RECHARGED.

ELECTRICAL ISOLATION -  FREQUENCY INPUTS AND OUTPUTS ARE ISOLATED FROM EACH OTHER AND FROM ALL OTHER CIRCUITS.  MULTIMETER IS ISOLATED FROM ALL OTHER CIRCUITS.  mV / TEMP, mA, AND BATTERY HAVE COMMON CONNECTIONS.

SIMULTANEOUS SIGNAL GENERATION - ALL SIGNAL CONNECTION, BOTH IN AND OUT, MAY BE USED SIMULTANEOUSLY EXCEPT WHERE PRECLUDED BY SWITCH SETTINGS. E.G. mV AND THERMOCOUPLE INDICATION ARE NOT AVAILABLE SIMULTANEOUSLY.  E.G. TWO FREQUENCY OUTPUTS, ONE THERMOCOUPLE SIMULATED SIGNAL, ANC mA OUT MAY ALL BE GENERATED SIMULTANEOUSLY.

    DISPLAYS
            TWO INDEPENDENT 5 DIGIT LED DISPLAYS
            TOP DISPLAY INDICATES mV OR TEMP. F. OR TEMP. C.
            BOTTOM DISPLAY INDICATES FREQUENCY CHANNEL 1 OR 2 OR AUXILIARY CHANNEL 3 USED
            WITH OPTIONAL MULTISOURCE MODULES.
            UPDATE RATE IS 1 PER SECOND
            TOP DISPLAY CYCLES ON AND OFF TO INDICATE OPEN THERMOCOUPLE CONDITION.

    PHYSICAL
            CASE - NEMA 12 PORTABLE DRAWN ALUMINUM ENCLOSURE
            SIZE -  10.25 in. HIGH, 11 in. WIDE, 7 in. DEEP
            WEIGHT - LESS THAN 12 lbs. WITHOUT ACCESSORIES

    ACCESSORIES
            CALIBRATION CERTIFICATE
            THREE 15 FOOT LONG CO-AXIAL TEST LEADS, TWO STACK ABLE BANANA ON ONE END AND
            TWO ALLIGATOR ON THE OTHER END.
            AC POWER CORD

  TOP